Design and rationale of the IN CONTROL trial: the effects of real-time continuous glucose monitoring on glycemia and quality of life in patients with type 1 diabetes mellitus and impaired awareness of hypoglycemia

Abstract

Background

Hypoglycemia is the main side effect of intensified insulin therapy in type 1 diabetes and recognized as a limitation in achieving glycemic targets. Patients with impaired awareness of hypoglycemia have a threefold to sixfold increased risk of severe hypoglycemia. Real-time continuous glucose monitoring may help patients with type 1 diabetes to achieve better glycemic control with less hypoglycemic episodes. Accordingly, one may hypothesize that particularly type 1 diabetes mellitus patients with impaired awareness of hypoglycemia will profit most from this technology with improvements in their quality of life. However, this has not yet been established. This trial aims to study the effect of real-time continuous glucose monitoring on glycemia and quality of life specifically in type 1 diabetes mellitus patients with established impaired awareness of hypoglycemia.

Methods/design

This is a two-center, randomized, cross-over trial with a 12-week wash-out period in between intervention periods. A total of 52 type 1 diabetes mellitus patients with impaired awareness of hypoglycemia according to Gold et al. criteria will be randomized to receive real-time continuous glucose monitoring or blinded continuous glucose monitoring for 16 weeks. After a wash-out period, patients will cross over to the other intervention. The primary outcome measure is time spent in euglycemia. Secondary outcomes include (diabetes-specific) markers of quality of life and other glycemic variables.

Discussion

It remains unclear whether patients with type 1 diabetes and impaired awareness of hypoglycemia benefit from real-time continuous glucose monitoring in real-life. This study will provide insight into the potential benefits of real-time continuous glucose monitoring in this patient population.

Trial registration

Clinicaltrials.gov: NCT01787903.

Peer Review reports

Background

Type 1 diabetes mellitus (T1DM) constitutes about 10-15 % of total diabetes rates and its incidence has been increasing worldwide at an alarming rate of 3-5 % per year [1]. T1DM is associated with an increased risk of microvascular and macrovascular co-morbidities. The Diabetes Control and Complication Trial/Epidemiology of Diabetes Interventions and Complications (DCCT/EDIC) studies have shown that intensive versus conventional glycemic control results in a reduction of microvascular [2] but also macrovascular complications [3]. However, the DCCT also demonstrated that intensive treatment associates with a considerably increased rate of hypoglycemia [2]. With incidences of approximately 2 per patient per week [46] for mild (i.e. self-treated) hypoglycemia, and 0.1-1.5 per patient year [2, 59] for severe hypoglycemia (SH), hypoglycemia is both the main limitation in achieving glycemic targets and the main side effect of intensified insulin therapy in T1DM [10, 11]. The American Diabetes Association (ADA) defines SH as an event requiring assistance of a third party, with symptoms of neuroglycopenia and recovery of neurological symptoms after restoration of plasma glucose [12].

Hypoglycemia is both a physical and psychological burden [13, 14]. It can interfere with every aspect of daily life, such as sleep, exercise, driving or otherwise travelling, but also social interactions and even employment [13]. Patients worry about having episodes of SH or getting the late complications of T1DM [6], and there is always the possibility of a hypoglycemic episode leading to a coma [6]. Recurrent hypoglycemia may promote the development of impaired awareness of hypoglycemia (IAH) by decreasing the glycemic threshold in the brain required for the activation of the autonomic system [15]. Consequently, by the time the glucose level is low enough to elicit symptoms, patients fail to recognize them due to neuroglycopenia. To date, no consensus on a satisfactory definition of IAH exists. We defined IAH as the diminished ability to recognize the onset of hypoglycemia [14]. Hypoglycemia awareness can be assessed by using self-rating questionnaires. The Gold method consists of one question: “do you know when hypoglycemia is commencing?”. The state of hypoglycemia awareness is then assessed by using a 7-point visual analogue scale, with 1 representing “always aware” and 7 representing “never aware”. A score of ≥4 suggests impaired awareness of hypoglycemia [16]. The Clarke method consists of eight questions identifying the respondents exposure to episodes of (mild and severe) hypoglycemia and examining the glycemic threshold for hypoglycemia. Impaired awareness of hypoglycemia is implied if the respondent has a score of ≥4 [17]. Both scoring systems show good performance in adults [18]. Impaired awareness of hypoglycemia renders patients at a threefold to sixfold increased risk of SH which considerably hampers their quality of life [13, 16, 19]. In addition, hypoglycemia can be fatal, with hypoglycemia mortality estimates ranging from 4 to 10 percent of deaths of patients with type 1 diabetes [20, 21]. Furthermore, a recent observational study showed that T1DM patients with an glycated hemoglobin level of 6.9 % or lower had a twice increased risk of death from any cause and cardiovascular causes compared with matched controls [22].

In 2006, real-time continuous glucose monitoring (RT-CGM) was introduced to assist patients in their self-management of blood glucose [23]. Real-time continuous glucose monitoring systems measure interstitial glucose levels and provide this information every five or ten minutes, with a delay of approximately 8 to 15 minutes [2426]. The added value lies in the display of trends and alarms that can be set to warn for impending hypo- or hyperglycemia [27, 28]. RT-CGM is associated with an improvement of glycemic control [27, 2941] and shorter duration of hypoglycemic episodes [4244]. However, the effect of conventional RT-CGM (i.e. without automated insulin suspension) specifically in patients with IAH has not yet been established in a real life setting [45]. In most studies either recent severe hypoglycemia was among the exclusion criteria [29] or the frequency of severe hypoglycemia at baseline was not mentioned [37, 40, 46]. Studies done in subjects with normal hypoglycemia awareness already indicate that use of RT-CGM might improve the impact of T1DM on daily life [28] and might increase treatment satisfaction [47], although an effect on the fear of hypoglycemia is not apparent [48]. Since experiencing the issues mentioned above might especially be the case for patients with IAH, it is not unlikely that use of RT-CGM might improve the quality of life in those patients.

This trial aims to study a wide range of effects of RT-CGM specifically in T1DM patients with established IAH, regardless of baseline HbA1c. We hypothesize that the use of RT-CGM, relative to a control intervention using masked CGM, will result in improvement of time spent in euglycemia and quality of life in T1DM patients with IAH.

Methods/design

Design

This is a two-center, randomized, cross-over trial with a 12-week wash-out period in between intervention periods (Fig. 1). Ethical approval has been granted by the medical ethical committee of the VU University Medical Center (VUMC).

Fig. 1
figure 1

General study plan

Recruitment

Subjects will be recruited from the outpatient clinic of the VUMC, Amsterdam and the Medical Center Haaglanden (MCH), The Hague, The Netherlands, from outpatient clinics at affiliated hospitals, but also from regional hospitals and via advertisements in local newspapers. Responders will be sent written extensive patient information. Written informed consent will be obtained from the subjects prior to any trial related procedure.

Study population

Fifty-two subjects with T1DM and IAH will be enrolled. Because patients with IAH often are already in good to moderate control in terms of HbA1c, patients will be included regardless of their HbA1c.

Inclusion criteria

  • T1DM, diagnosed according to ADA criteria [49] regardless duration.

  • Use of multiple daily injections of insulin or continuous subcutaneous insulin infusion. (Participants may not change their treatment method during the study.)

  • Age between 18 and 75 years (inclusive).

  • IAH according to the questionnaire by Gold et al. [16].

  • Performing self-monitored blood glucose measurements (SMBG) at least 3/day or 21/week.

Exclusion criteria

  • History of (recent) major renal, liver, or (ischemic) heart disease (including cardiac conduction disorders).

  • Current untreated proliferative diabetic retinopathy.

  • Current (treatment for) malignancy.

  • Current use of non-selective beta-blockers.

  • Current psychiatric disorders, including schizophrenia, bipolar disorder, anorexia nervosa or bulimia nervosa.

  • Substance abuse or alcohol abuse (men >21 units/week, women >14 units/week).

  • Current pregnancy or intention to conceive.

  • Current use of RT-CGM other than for short term (i.e. diagnostic use or use shorter than 3 consecutive months).

  • Hearing or vision impairment hindering perceiving of glucose display and alarms, or otherwise incapable of using a (RT-)CGM, in the opinion of the investigator.

  • Poor commandment of the Dutch language or any (mental) disorder that precludes full understanding of the purpose and instructions of the study.

  • Participation in another clinical study.

  • Known or suspected allergy to trial product or related products.

Study plan

A detailed overview of study visits, telephone consultations and procedures is given in Additional file 1. At each visit, a recent history will be taken, including e.g. complaints, adverse events, changes in medication treatment regime and the occurrence of SH (i.e. requiring assistance of a third party). Weight and vital signs will be measured at each visit. Outcome measurements will be performed at baseline and after each 16-week intervention phase (Fig. 1).

Screening

Hypoglycemia awareness will be assessed using the questionnaire developed by Gold et al. and by Clarke et al. [16, 17]. The frequency and severity of severe hypoglycemia will be assessed using the self-report questionnaire according to Langan et al. [50]. Subsequently, a complete medical and socio-economic history will be taken and a complete physical examination will be performed. Participants will be examined for the presence of diabetes-related (microvascular) complications. Thus, plasma creatinine, calculated estimated glomerular filtration rate (eGFR) and albumin-creatinine ratio (ACR) in a random urine sample will be measured to evaluate diabetic nephropathy [51, 52]. The presence/severity of diabetic sensorimotor polyneuropathy will be quantified by the Modified Toronto Clinical Scoring System [53]. For the assessment of vibration sense, the vibration perception threshold will be quantified using a neurothesiometer [54]. The presence/severity of diabetic retinopathy (DRP) will be assessed based on the relevant medical history and a recent (<6 months) evaluation by the patient’s ophthalmologist. In the absence of this information, patients will be either requested to visit their ophthalmologist prior to study onset or they will be referred for fundus photography [55]. Blood will be collected to ascertain in- and exclusion criteria and for baseline measurement.

Run-in phase

A 5-week run-in phase, starting after inclusion, will be used to allow for study effects and enable diabetes- and study-related education. All participants will receive approximately 30 minutes of individual education in the principles of diabetes management, including basic principles of standardized SMBG, glucose fluctuations, insulin and carbohydrates, hyper- and hypoglycemia, impaired awareness of hypoglycemia and RT-CGM. In case a participant does not use the technique of carbohydrate counting, no education on this subject will be given in order to prevent confounding. Participants will be equipped with a masked CGM device to be worn for two weeks to gather baseline data and to familiarize them with longer-term wearing of this type of device. During follow-up visits, there will be an ongoing evaluation of diabetes self-management and knowledge of diabetes management, as this is part of standard diabetes care. A test of knowledge will not be used.

Intervention phase

Randomization

After successful participation in the run-in phase, participants will be randomized, using block randomization (allocation 1:1), to the first intervention period. A sealed envelope for each participant will be drawn and opened by qualified study staff. In addition, the block size will be determined by the institutional trial pharmacist. Study staff and participants will not be blinded for treatment order, as this is an open intervention study. When allocated to RT-CGM, participants will receive additional education on how to use RT-CGM and the accompanying specialized software. Participants will be asked to upload their RT-CGM-data before each follow-up visit. When allocated to masked CGM, participants will continue to wear their CGM device from the run-in phase. Participants are encouraged to wear the assigned study devices continuously.

Follow-up visits and telephone consultations

During the intervention periods there will be monthly follow-up visits and telephone consultations involving inquiry after e.g. actual complaints/symptoms, episodes of (severe) hypoglycemia, study device use and related technical issues, and medication using a checklist. During both intervention periods, research physicians will guide participants in concordance with the ADA Standards of Medical Care in Diabetes [56]. The guidance will be equal during both intervention periods. During the visits, therapy adjustments will be discussed and logged, based on RT-CGM-data in the RT-CGM group or SMBG-data in the masked CGM group. No treatment or insulin titration protocol will be used, neither will SMBG use be standardized, in order to avoid additional interventions.

Wash-out phase

After the first intervention phase, participants will enter a 12-week wash-out phase, during which they will only receive two-weekly telephone consultations for taking recent histories and monitoring of potential adverse events. At the end of the wash-out period, general diabetes and CGM education will be given similar to the first run-in phase. Also, participants will start to wear masked CGM again for two weeks to gather baseline data for the second intervention period.

Endpoints

Primary endpoint

The mean difference in time spent in euglycemia (interstitial glucose range >3.9 mmol/L - ≤10.0 mmol/L), expressed as hours/day between the two intervention periods. The specific ranges of euglycemia are based on the definition of the ADA and comparing literature [12, 44].

Secondary endpoints

  • (Diabetes-specific) markers of quality of life, as assessed with questionnaires covering: diabetes-related emotional distress (PAID-5 (Cronbach’s α = 0.86) [57]), fear of hypoglycemia (HFS-2 (Cronbach’s α = 0.94) [58, 59]), diabetes self-efficacy (CIDS (Cronbach’s α = 0.94) [60]), generic health-status (EQ5D (Cronbach’s α = 0.73) [61, 62]) and emotional well-being (Cronbach’s α = 0.82) (WHO-5 [6365]).

  • Other glycemic variables, including HbA1c, time spent in hypo- and hyperglycemia ranges, frequency of severe (i.e. requiring assistance of a third party) and mild (i.e. self-treated) hypoglycemia (assessed by analyzing (RT-)CGM data) and duration of hypoglycemia.

  • Changes in hypoglycemia awareness score according to Gold et al. [16].

  • Glycemic variability. Glucose variability (defined as inter-day and intraday glycemic variability), will be calculated as Mean Of Daily Differences and Continuous Overall Net Glycemic Action, using all (RT-)CGM values during the 16 week intervention period [66, 67].

Exploratory endpoints

  • Autonomic nervous system function. The functioning of the autonomic nervous system will be indirectly assessed by analysing heart rate variability and blood pressure changes during four of the five Ewing’s standardized cardiovascular reflex tests: the Valsalva ratio, the 30:15 ratio on standing up, the maximum-minimum heart rate during deep breathing and postural blood pressure change [6871]. The heart-rate variability will be assessed using non-invasive, automated beat-to-beat blood pressure and ECG recordings (Nexfin®, BM Eye, Amsterdam, the Netherlands). that will be analyzed by dedicated software, automatically calculating HRV indices for both the time-domain and frequency-domain variables. Quality control will be performed by research staff by visual inspection of the data. Data will be excluded from analysis if >5 % of the measured beats are extrasystoles. No agreement exists on the number of abnormal cardiovascular tests required to reach the diagnosis of cardiovascular autonomic neuropathy [71]. However, an abnormality of more than one test, on more than one occasion, is indicative of (cardiovascular) autonomic dysfunction [71, 72].

  • Duration of sensor wear.

  • Changes in hypoglycemia awareness score according to Clarke et al. [17].

  • Satisfaction with use of RT-CGM, assessed by the CGM-SAT questionnaire [73].

Study devices

The MiniMed Paradigm® Veo™ System (Medtronic, Northridge, CA) will be used as RT-CGM device. The system consists of a continuous Enlite™ glucose sensor, a MiniLink transmitter and the Paradigm® Veo™ System [74, 75]. The sensor is a membrane-covered enzyme coated electrode placed through the skin into the subcutaneous space using an auto-insertion device. This sensor has to be changed every 6 days . The wireless MiniLinktransmitter is a small rechargeable device that is connected to the glucose sensor and sends glucose data wirelessly to the monitor every 5 minutes, 24 hours a day. The monitor shows real-time glucose measurements and provides information regarding changes in glucose levels (trends). Calibration is required every 12 hours. The system allows to set alarms, i.e. alarms for low and high glucose limits, as agreed with the health-care provider, at which the system should alert the patient whenever glucose values are approximating or exceeding these pre-set values. The low-glucose limit during this trial will be pre-set at 4,5 mmol/L and cannot be lowered in order to prevent hypoglycemia. Also, accuracy of glucose sensors decreases in hypoglycemic ranges [76, 77]. Low-glucose suspension function will not be used. Finally, the system shows participants their real-time glucose values and patterns, and allows for adjustment of treatment and/or lifestyle accordingly.

The masked CGM device that will be used is the iPro™2 Continuous Glucose Monitor (Medtronic, Northridge, CA), which also uses the Enlite™ glucose sensor [74, 75]. Masked CGM- and SMBG-data must be uploaded every week for the purpose of (retrospective) calibration of the iPro™2. Participants will be blinded to the CGM-data. The investigators will review the CGM-data for quality based on duration of sensor wear, number of sensor values, accuracy (mean absolute difference), number of valid calibrations. In case of missing data due to low quality CMG-data, the intervention phase will be extended.

Statistical considerations

Analysis

Intention to treat analysis will include all randomized subjects, including drop-outs. Per-protocol analysis will include only subjects who have completed at least the first four weeks of the first intervention period. The total evaluable sample will consist of all intention-to-treat subjects who complete both intervention periods according to the protocol. Evaluation of (RT-)CGM data will not be performed in an assessor-blinded way. Quality control of the (RT-)CGM data and data analysis will be performed by the same study staff.

Unless otherwise stated, all statistical tests will be conducted at a 2-tailed significance level of 0.05. According to their distribution, the various parameters will be expressed as mean (± SD), median (interquartile range) or number (%). Residual effects will be ascertained before further analysis will be performed using non-parametric tests [78, 79]. The primary endpoint, i.e. mean difference in time spent in the euglycemic range, expressed as hours/day, during the RT-CGM versus the masked CGM phases, will be analyzed per month using mixed model analysis, in order to show a possible effect over time. In the mixed model analysis, time spent in euglycemia will be used as dependent variable. Independent variables will include: treatment arm, interaction between RT-CGM and CGM, and time in months since start of the intervention. Baseline CGM data, gathered during the run-in phase, will be implemented as covariate. In case of carry-over effects and an unequal distribution of treatment method (MDI / CSII) over the intervention orders (RT-CGM – CGM / CGM – RT-CGM), treatment method will also be included as covariate. Post hoc tests will be corrected using Bonferroni correction. If additional testing will reveal potential confounders, adjustments for these factors may necessitate analysis of covariance (ANCOVA). By using mixed model analysis, unequal use of the devices will not necessitate correction, as long as missing data can be considered to be at random.

Carry-over

To date, little is known about the possible carry-over effects of RT-CGM. Yet such an effect would seem logical considering the expected educational effect of RT-CGM in e.g. the ability of participants to recognize patterns in otherwise undetected hypo- or hyperglycemic events and adjust insulin therapy accordingly. To minimize carry-over effects, a washout period of 12 weeks will be implanted. We consider a 12-week period sufficient for the behavioral modification component to wear off. Additionally, a 12-week time period allows realistic HbA1c changes resulting from patients self-management during that same period to establish a reliable baseline for the second intervention period.

Sample size calculation

Since current RT-CGM technology may show relatively poor performance in the hypoglycemic range [80], we chose mean time (hours/day) spent in euglycemia, i.e. glucose values ranging >3.9 - ≤10.0 mmol/L, during the study period as our primary end-point. A previously published intervention study using RT-CGM demonstrated a difference of 1.5 hours in time spent in euglycemia between the RT-CGM and control group [44]. In order to detect a difference of 1.5 hours (6.25 % from 24 hours) in time spent in euglycemia, assuming a standard deviation of 3.5, alpha-level of 0.05, power of 80 %, a drop-out rate of 15 % and a correlation of 0.5 between repeated measures (which is due to the cross-over design of the study in which subjects serve as their own control), a sample size of 52 patients will be needed.

Discussion

Although it seems obvious that T1DM patients with IAH could potentially benefit from RT-CGM, the effect of conventional RT-CGM (without automated insulin suspension) has not yet been established in an ambulatory setting. A hyperinsulinemic hypoglycemic clamp study of Ly et al. showed that 4 weeks of RT-CGM improved epinephrine responses in young T1DM patients with IAH, suggesting that IAH can be restored in adolescents by using RT-CGM [81]. Furthermore, the first observational study performed in patients with IAH demonstrated a clear reduction of SH with RT-CGM use [82], addressing the need for further intervention studies in patients with IAH. A randomized controlled trial of Little et al. demonstrated that IAH can be restored and SH prevented using existing technology [45]. In contrast to expectations, RT-CGM was not superior over self-monitoring of blood glucose by finger prick. However, during this trial, strict insulin titration protocols where used, which may not reflect real-life diabetes management. Although Ly et al. showed that sensor-augmented insulin pump therapy with automated insulin suspension reduced the frequency of SH significantly in T1DM patients with IAH [83], this reduction lost significance when 2 outliers, whose rates of hypoglycemia were higher at baseline, were removed from analysis. Hence, it remains unclear whether T1DM patients with IAH benefit from conventional RT-CGM in real-life.

The study protocol poses some limitations. Hypoglycemia awareness will be assessed using the subjective self-rating questionnaire of Gold et al. [16]. Objective assessments of hypoglycemia awareness, i.e. by using CGM data or by inducing hypoglycemia in a laboratory setting, will not be used. Self-reported hypoglycemia awareness has shown to be reliable in previous studies [1618, 84] and reflects clinical practice. Also, continuous use of RT-CGM is encouraged, but not mandatory, which could influence the effect of RT-CGM on glycemia [40]. Furthermore, the low alarm value used in this trial is pre-set at 4,5 mmol/L in order to prevent hypoglycemia. In clinical practice however, alarm values should be evaluated and set individually. Moreover, the decrease in sensor accuracy in hypoglycemic ranges could compromise the analysis of glycemia [76, 77]. Finally, participants already using an insulin pump are offered a second device, which could diminish compliance with regards to logging SMBG values, meals, insulin and activities. This could affect the value of interpreting the RT-CGM-data retrospectively.

With this study, we expect to gain a better understanding of the potential benefits of RT-CGM technology in a sample of patients with T1DM complicated by impaired hypoglycemia awareness. Also, the study is likely to yield clinically relevant information to help further improve targeted intervention strategies to help these patients improve their diabetes self-management and quality of life.

Abbreviations

ADA:

American Diabetes Association

ACR:

Albumin-creatinine ratio

CGM:

Continuous glucose monitoring

CGM-SAT:

Questionnaire for satisfaction with use of CGM

CIDS:

Confidence in diabetes self-management

DCCT/EDIC:

The diabetes control and complication trial/epidemiology of diabetes interventions and complications

DRP:

Diabetic retinopathy

EQ-5D:

EuroQol-5D

eGFR:

estimated glomerular filtration rate

HbA1c:

Glycosylated hemoglobin

HFS:

Hypoglycemia fear scale

IAH:

Impaired awareness of hypoglycemia

MCH:

Medical Center Haaglanden

PAID-5:

Problem areas in diabetes scale 5-items

QoL:

Quality of life

RT-CGM:

Real-time continuous glucose monitoring

SH:

Severe hypoglycemia

SMBG:

Self-monitoring of blood glucose

T1DM:

Type 1 diabetes mellitus

VUMC:

VU University Medical Center

WHO-5:

World Health Organization-Five Well-Being Index

References

  1. Forlenza GP, Rewers M. The epidemic of type 1 diabetes: what is it telling us? Curr Opin Endocrinol Diabetes Obes. 2011;18:248–51.

    Article  PubMed  Google Scholar 

  2. The Diabetes Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. The Diabetes Control and Complications Trial Research Group. N Engl J Med. 1993;329:977–86.

    Article  Google Scholar 

  3. Nathan DM, Cleary PA, Backlund JY, Genuth SM, Lachin JM, Orchard TJ, et al. Intensive diabetes treatment and cardiovascular disease in patients with type 1 diabetes. N Engl J Med. 2005;353:2643–53.

    Article  PubMed  Google Scholar 

  4. MacLeod KM, Gold AE, Frier BM. Frequency, severity and symptomatology of hypoglycaemia: a comparative trial of human and porcine insulins in type 1 diabetic patients. Diabet Med. 1995;12:134–41.

    Article  CAS  PubMed  Google Scholar 

  5. Pedersen-Bjergaard U, Pramming S, Heller SR, Wallace TM, Rasmussen AK, Jorgensen HV, et al. Severe hypoglycaemia in 1076 adult patients with type 1 diabetes: influence of risk markers and selection. Diabetes Metab Res Rev. 2004;20:479–86.

    Article  PubMed  Google Scholar 

  6. Pramming S, Thorsteinsson B, Bendtson I, Binder C. Symptomatic hypoglycaemia in 411 type 1 diabetic patients. Diabet Med. 1991;8:217–22.

    Article  CAS  PubMed  Google Scholar 

  7. Donnelly LA, Morris AD, Frier BM, Ellis JD, Donnan PT, Durrant R, et al. Frequency and predictors of hypoglycaemia in Type 1 and insulin-treated Type 2 diabetes: a population-based study. Diabet Med. 2005;22:749–55.

    Article  CAS  PubMed  Google Scholar 

  8. MacLeod KM, Hepburn DA, Frier BM. Frequency and morbidity of severe hypoglycaemia in insulin-treated diabetic patients. Diabet Med. 1993;10:238–45.

    Article  CAS  PubMed  Google Scholar 

  9. ter Braak EW, Appelman AM, van de Laak M, Stolk RP, van Haeften TW, Erkelens DW. Clinical characteristics of type 1 diabetic patients with and without severe hypoglycemia. Diabetes Care. 2000;23:1467–71.

    Article  PubMed  Google Scholar 

  10. Cryer PE. Hypoglycaemia: the limiting factor in the glycaemic management of Type I and Type II diabetes. Diabetologia. 2002;45:937–48.

    Article  CAS  PubMed  Google Scholar 

  11. Cryer PE, Davis SN, Shamoon H. Hypoglycemia in diabetes. Diabetes Care. 2003;26:1902–12.

    Article  CAS  PubMed  Google Scholar 

  12. Workgroup on Hypoglycemia ADA. Defining and reporting hypoglycemia in diabetes: a report from the American Diabetes Association Workgroup on Hypoglycemia. Diabetes Care. 2005;28:1245–9.

    Article  Google Scholar 

  13. Frier BM. How hypoglycaemia can affect the life of a person with diabetes. Diabetes Metab Res Rev. 2008;24:87–92.

    Article  PubMed  Google Scholar 

  14. Frier BM. Hypoglycaemia in diabetes mellitus: epidemiology and clinical implications. Nat Rev Endocrinol. 2014;10(12):711–22.

    Article  CAS  PubMed  Google Scholar 

  15. Hepburn DA, Patrick AW, Brash HM, Thomson I, Frier BM. Hypoglycaemia unawareness in type 1 diabetes: a lower plasma glucose is required to stimulate sympatho-adrenal activation. Diabet Med. 1991;8:934–45.

    Article  CAS  PubMed  Google Scholar 

  16. Gold AE, MacLeod KM, Frier BM. Frequency of severe hypoglycemia in patients with type I diabetes with impaired awareness of hypoglycemia. Diabetes Care. 1994;17:697–703.

    Article  CAS  PubMed  Google Scholar 

  17. Clarke WL, Cox DJ, Gonder-Frederick LA, Julian D, Schlundt D, Polonsky W. Reduced awareness of hypoglycemia in adults with IDDM. A prospective study of hypoglycemic frequency and associated symptoms. Diabetes Care. 1995;18:517–22.

    Article  CAS  PubMed  Google Scholar 

  18. Geddes J, Wright RJ, Zammitt NN, Deary IJ, Frier BM. An evaluation of methods of assessing impaired awareness of hypoglycemia in type 1 diabetes. Diabetes Care. 2007;30:1868–70.

    Article  CAS  PubMed  Google Scholar 

  19. Geddes J, Schopman JE, Zammitt NN, Frier BM. Prevalence of impaired awareness of hypoglycaemia in adults with Type 1 diabetes. Diabet Med. 2008;25:501–4.

    Article  CAS  PubMed  Google Scholar 

  20. Feltbower RG, Bodansky HJ, Patterson CC, Parslow RC, Stephenson CR, Reynolds C, et al. Acute complications and drug misuse are important causes of death for children and young adults with type 1 diabetes: results from the Yorkshire Register of diabetes in children and young adults. Diabetes Care. 2008;31:922–6.

    Article  PubMed  Google Scholar 

  21. Patterson CC, Dahlquist G, Harjutsalo V, Joner G, Feltbower RG, Svensson J, et al. Early mortality in EURODIAB population-based cohorts of type 1 diabetes diagnosed in childhood since 1989. Diabetologia. 2007;50:2439–42.

    Article  CAS  PubMed  Google Scholar 

  22. Lind M, Svensson AM, Kosiborod M, Gudbjornsdottir S, Pivodic A, Wedel H, et al. Glycemic control and excess mortality in type 1 diabetes. N Engl J Med. 2014;371:1972–82.

    Article  PubMed  Google Scholar 

  23. Garg S, Zisser H, Schwartz S, Bailey T, Kaplan R, Ellis S, et al. Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor: a randomized controlled trial. Diabetes Care. 2006;29:44–50.

    Article  CAS  PubMed  Google Scholar 

  24. Feldman B, Brazg R, Schwartz S, Weinstein R. A continuous glucose sensor based on wired enzyme technology -- results from a 3-day trial in patients with type 1 diabetes. Diabetes Technol Ther. 2003;5:769–79.

    Article  CAS  PubMed  Google Scholar 

  25. Keenan DB, Mastrototaro JJ, Voskanyan G, Steil GM. Delays in minimally invasive continuous glucose monitoring devices: a review of current technology. J Diabetes Sci Technol. 2009;3:1207–14.

    Article  PubMed  PubMed Central  Google Scholar 

  26. Weinstein RL, Schwartz SL, Brazg RL, Bugler JR, Peyser TA, McGarraugh GV. Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements. Diabetes Care. 2007;30:1125–30.

    Article  CAS  PubMed  Google Scholar 

  27. Bode B, Gross K, Rikalo N, Schwartz S, Wahl T, Page C, et al. Alarms based on real-time sensor glucose values alert patients to hypo- and hyperglycemia: the guardian continuous monitoring system. Diabetes Technol Ther. 2004;6:105–13.

    Article  CAS  PubMed  Google Scholar 

  28. Geelhoed-Duijvestijn PH, Box F, Diamant M, Otoide-Vree M, Niel Van J. Real time continuous glucose monitoring: how does it work in real life? Br J Diab Vasc Dis. 2011;11:145–9.

    Article  Google Scholar 

  29. Bergenstal RM, Tamborlane WV, Ahmann A, Buse JB, Dailey G, Davis SN, et al. Effectiveness of sensor-augmented insulin-pump therapy in type 1 diabetes. N Engl J Med. 2010;363:311–20.

    Article  CAS  PubMed  Google Scholar 

  30. Bergenstal RM, Tamborlane WV, Ahmann A, Buse JB, Dailey G, Davis SN, et al. Sensor-augmented pump therapy for A1C reduction (STAR 3) study: results from the 6-month continuation phase. Diabetes Care. 2011;34:2403–5.

    Article  PubMed  PubMed Central  Google Scholar 

  31. Bode B, Beck RW, Xing D, Gilliam L, Hirsch I, Kollman C, et al. Sustained benefit of continuous glucose monitoring on A1C, glucose profiles, and hypoglycemia in adults with type 1 diabetes. Diabetes Care. 2009;32:2047–9.

    Article  PubMed  Google Scholar 

  32. Danne T, De Valk HW, Kracht T, Walte K, Geldmacher R, Solter L, et al. Reducing glycaemic variability in type 1 diabetes self-management with a continuous glucose monitoring system based on wired enzyme technology. Diabetologia. 2009;52:1496–503.

    Article  CAS  PubMed  Google Scholar 

  33. Deiss D, Bolinder J, Riveline JP, Battelino T, Bosi E, Tubiana-Rufi N, et al. Improved glycemic control in poorly controlled patients with type 1 diabetes using real-time continuous glucose monitoring. Diabetes Care. 2006;29:2730–2.

    Article  PubMed  Google Scholar 

  34. Garg SK, Schwartz S, Edelman SV. Improved glucose excursions using an implantable real-time continuous glucose sensor in adults with type 1 diabetes. Diabetes Care. 2004;27:734–8.

    Article  PubMed  Google Scholar 

  35. Hirsch IB, Abelseth J, Bode BW, Fischer JS, Kaufman FR, Mastrototaro J, et al. Sensor-augmented insulin pump therapy: results of the first randomized treat-to-target study. Diabetes Technol Ther. 2008;10:377–83.

    Article  CAS  PubMed  Google Scholar 

  36. Ludvigsson J, Hanas R. Continuous subcutaneous glucose monitoring improved metabolic control in pediatric patients with type 1 diabetes: a controlled crossover study. Pediatrics. 2003;111:933–8.

    Article  PubMed  Google Scholar 

  37. Raccah D, Sulmont V, Reznik Y, Guerci B, Renard E, Hanaire H, et al. Incremental value of continuous glucose monitoring when starting pump therapy in patients with poorly controlled type 1 diabetes: the RealTrend study. Diabetes Care. 2009;32:2245–50.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  38. Radermecker RP, Saint RA, Scheen AJ, Bringer J, Renard E. Continuous glucose monitoring reduces both hypoglycaemia and HbA1c in hypoglycaemia-prone type 1 diabetic patients treated with a portable pump. Diabetes Metab. 2010;36:409–13.

    Article  CAS  PubMed  Google Scholar 

  39. Rodbard D, Bailey T, Jovanovic L, Zisser H, Kaplan R, Garg SK. Improved quality of glycemic control and reduced glycemic variability with use of continuous glucose monitoring. Diabetes Technol Ther. 2009;11:717–23.

    Article  CAS  PubMed  Google Scholar 

  40. Tamborlane WV, Beck RW, Bode BW, Buckingham B, Chase HP, Clemons R, et al. Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med. 2008;359:1464–76.

    Article  PubMed  Google Scholar 

  41. Hermanides J, Norgaard K, Bruttomesso D, Mathieu C, Frid A, Dayan CM, et al. Sensor-augmented pump therapy lowers HbA(1c) in suboptimally controlled Type 1 diabetes; a randomized controlled trial. Diabet Med. 2011;28:1158–67.

    Article  CAS  PubMed  Google Scholar 

  42. Davey RJ, Jones TW, Fournier PA. Effect of short-term use of a continuous glucose monitoring system with a real-time glucose display and a low glucose alarm on incidence and duration of hypoglycemia in a home setting in type 1 diabetes mellitus. J Diabetes Sci Technol. 2010;4:1457–64.

    Article  PubMed  PubMed Central  Google Scholar 

  43. Beck RW, Hirsch IB, Laffel L, Tamborlane WV, Bode BW, Buckingham B, et al. The effect of continuous glucose monitoring in well-controlled type 1 diabetes. Diabetes Care. 2009;32:1378–83.

    Article  CAS  PubMed  Google Scholar 

  44. Battelino T, Phillip M, Bratina N, Nimri R, Oskarsson P, Bolinder J. Effect of continuous glucose monitoring on hypoglycemia in type 1 diabetes. Diabetes Care. 2011;34:795–800.

    Article  PubMed  PubMed Central  Google Scholar 

  45. Little SA, Leelarathna L, Walkinshaw E, Tan HK, Chapple O, Lubina-Solomon A, et al. Recovery of Hypoglycemia Awareness in Long-standing Type 1 Diabetes: A Multicenter 2 x 2 Factorial Randomized Controlled Trial Comparing Insulin Pump With Multiple Daily Injections and Continuous With Conventional Glucose Self-monitoring (HypoCOMPaSS). Diabetes Care. 2014;37:2114–22.

    Article  CAS  PubMed  Google Scholar 

  46. Deiss D, Hartmann R, Schmidt J, Kordonouri O. Results of a randomised controlled cross-over trial on the effect of continuous subcutaneous glucose monitoring (CGMS) on glycaemic control in children and adolescents with type 1 diabetes. Exp Clin Endocrinol Diabetes. 2006;114:63–7.

    Article  CAS  PubMed  Google Scholar 

  47. Tansey M, Laffel L, Cheng J, Beck R, Coffey J, Huang E, et al. Satisfaction with continuous glucose monitoring in adults and youths with Type 1 diabetes. Diabet Med. 2011;28:1118–22.

    Article  CAS  PubMed  Google Scholar 

  48. Davey RJ, Stevens K, Jones TW, Fournier PA. The effect of short-term use of the Guardian RT continuous glucose monitoring system on fear of hypoglycaemia in patients with type 1 diabetes mellitus. Prim Care Diabetes. 2011;6(1):35–9.

    Article  PubMed  Google Scholar 

  49. American Diabetes Association. Diagnosis and classification of diabetes mellitus. Diabetes Care. 2012;35(1):S64–71.

    Article  Google Scholar 

  50. Langan SJ, Deary IJ, Hepburn DA, Frier BM. Cumulative cognitive impairment following recurrent severe hypoglycaemia in adult patients with insulin-treated diabetes mellitus. Diabetologia. 1991;34:337–44.

    Article  CAS  PubMed  Google Scholar 

  51. Nathan DM, Rosenbaum C, Protasowicki VD. Single-void urine samples can be used to estimate quantitative microalbuminuria. Diabetes Care. 1987;10:414–8.

    Article  CAS  PubMed  Google Scholar 

  52. Mogensen CE, Keane WF, Bennett PH, Jerums G, Parving HH, Passa P, et al. Prevention of diabetic renal disease with special reference to microalbuminuria. Lancet. 1995;346:1080–4.

    Article  CAS  PubMed  Google Scholar 

  53. Bril V, Tomioka S, Buchanan RA, Perkins BA. Reliability and validity of the modified Toronto Clinical Neuropathy Score in diabetic sensorimotor polyneuropathy. Diabet Med. 2009;26:240–6.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  54. Martin CL, Waberski BH, Pop-Busui R, Cleary PA, Catton S, Albers JW, et al. Vibration perception threshold as a measure of distal symmetrical peripheral neuropathy in type 1 diabetes: results from the DCCT/EDIC study. Diabetes Care. 2010;33:2635–41.

    Article  PubMed  PubMed Central  Google Scholar 

  55. Aldington SJ, Kohner EM, Meuer S, Klein R, Sjolie AK. Methodology for retinal photography and assessment of diabetic retinopathy: the EURODIAB IDDM complications study. Diabetologia. 1995;38:437–44.

    Article  CAS  PubMed  Google Scholar 

  56. Guidelines ADA. Standards of medical care in diabetes--2012. Diabetes Care. 2012;2012(35 Suppl 1):S11–63.

    Google Scholar 

  57. McGuire BE, Morrison TG, Hermanns N, Skovlund S, Eldrup E, Gagliardino J, et al. Short-form measures of diabetes-related emotional distress: the Problem Areas in Diabetes Scale (PAID)-5 and PAID-1. Diabetologia. 2010;53:66–9.

    Article  CAS  PubMed  Google Scholar 

  58. Cox DJ, Irvine A, Gonder-Frederick L, Nowacek G, Butterfield J. Fear of hypoglycemia: quantification, validation, and utilization. Diabetes Care. 1987;10:617–21.

    Article  CAS  PubMed  Google Scholar 

  59. Gonder-Frederick LA, Schmidt KM, Vajda KA, Greear ML, Singh H, Shepard JA, et al. Psychometric properties of the hypoglycemia fear survey-ii for adults with type 1 diabetes. Diabetes Care. 2011;34:801–6.

    Article  PubMed  PubMed Central  Google Scholar 

  60. Van Der Ven NC, Weinger K, Yi J, Pouwer F, Ader H, Van Der Ploeg HM, et al. The confidence in diabetes self-care scale: psychometric properties of a new measure of diabetes-specific self-efficacy in Dutch and US patients with type 1 diabetes. Diabetes Care. 2003;26:713–8.

    Article  PubMed Central  Google Scholar 

  61. Linde L, Sorensen J, Ostergaard M, Horslev-Petersen K, Hetland ML. Health-related quality of life: validity, reliability, and responsiveness of SF-36, 15D, EQ-5D [corrected] RAQoL, and HAQ in patients with rheumatoid arthritis. J Rheumatol. 2008;35:1528–37.

    PubMed  Google Scholar 

  62. Savoia E, Fantini MP, Pandolfi PP, Dallolio L, Collina N. Assessing the construct validity of the Italian version of the EQ-5D: preliminary results from a cross-sectional study in North Italy. Health Qual Life Outcomes. 2006;4:47.

    Article  PubMed  PubMed Central  Google Scholar 

  63. Awata S, Bech P, Yoshida S, Hirai M, Suzuki S, Yamashita M, et al. Reliability and validity of the Japanese version of the World Health Organization-Five Well-Being Index in the context of detecting depression in diabetic patients. Psychiatry Clin Neurosci. 2007;61:112–9.

    Article  PubMed  Google Scholar 

  64. De Wit M, Pouwer F, Gemke RJ, Delemarre-van De Waal HA, Snoek FJ. Validation of the WHO-5 Well-Being Index in adolescents with type 1 diabetes. Diabetes Care. 2007;30:2003–6.

    Article  PubMed  Google Scholar 

  65. Lowe B, Spitzer RL, Grafe K, Kroenke K, Quenter A, Zipfel S, et al. Comparative validity of three screening questionnaires for DSM-IV depressive disorders and physicians’ diagnoses. J Affect Disord. 2004;78:131–40.

    Article  PubMed  Google Scholar 

  66. McDonnell CM, Donath SM, Vidmar SI, Werther GA, Cameron FJ. A novel approach to continuous glucose analysis utilizing glycemic variation. Diabetes Technol Ther. 2005;7:253–63.

    Article  CAS  PubMed  Google Scholar 

  67. Molnar GD, Taylor WF, Ho MM. Day-to-day variation of continuously monitored glycaemia: a further measure of diabetic instability. Diabetologia. 1972;8:342–8.

    Article  CAS  PubMed  Google Scholar 

  68. Jarrin DC, McGrath JJ, Giovanniello S, Poirier P, Lambert M. Measurement fidelity of heart rate variability signal processing: The devil is in the details. Int J Psychophysiol. 2012;86(1):88–97.

    Article  PubMed  Google Scholar 

  69. Spallone V, Ziegler D, Freeman R, Bernardi L, Frontoni S, Pop-Busui R, et al. Cardiovascular autonomic neuropathy in diabetes: clinical impact, assessment, diagnosis, and management. Diabetes Metab Res Rev. 2011;27(7):639–53.

    Article  PubMed  Google Scholar 

  70. Ewing DJ, Martyn CN, Young RJ, Clarke BF. The value of cardiovascular autonomic function tests: 10 years experience in diabetes. Diabetes Care. 1985;8:491–8.

    Article  CAS  PubMed  Google Scholar 

  71. Spallone V, Bellavere F, Scionti L, Maule S, Quadri R, Bax G, et al. Recommendations for the use of cardiovascular tests in diagnosing diabetic autonomic neuropathy. Nutr Metab Cardiovasc Dis. 2011;21:69–78.

    Article  CAS  PubMed  Google Scholar 

  72. Consensus statement: Report and recommendations of the San Antonio conference on diabetic neuropathy. American Diabetes Association American Academy of Neurology. Diabetes Care 1988, 11: 592–597.

  73. Diabetes Research in Children Network (DirecNet) Study Group. Youth and parent satisfaction with clinical use of the GlucoWatch G2 Biographer in the management of pediatric type 1 diabetes. Diabetes Care. 2005;28:1929–35.

    Article  Google Scholar 

  74. Kropff J, Bruttomesso D, Doll W, Farret A, Galasso S, Luijf YM, et al. Accuracy of two continuous glucose monitoring systems: a head-to-head comparison under clinical research centre and daily life conditions. Diabetes Obes Metab. 2015;17:343–9.

    Article  CAS  PubMed  Google Scholar 

  75. Keenan DB, Cartaya R, Mastrototaro JJ. Accuracy of a new real-time continuous glucose monitoring algorithm. J Diabetes Sci Technol. 2010;4:111–8.

    Article  PubMed  PubMed Central  Google Scholar 

  76. Clarke WL, Anderson S, Farhy L, Breton M, Gonder-Frederick L, Cox D, et al. Evaluating the clinical accuracy of two continuous glucose sensors using continuous glucose-error grid analysis. Diabetes Care. 2005;28:2412–7.

    Article  CAS  PubMed  Google Scholar 

  77. Kovatchev B, Anderson S, Heinemann L, Clarke W. Comparison of the numerical and clinical accuracy of four continuous glucose monitors. Diabetes Care. 2008;31:1160–4.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  78. Brown Jr BW. The crossover experiment for clinical trials. Biometrics. 1980;36:69–79.

    Article  PubMed  Google Scholar 

  79. Koch GG. The use of non-parametric methods in the statistical analysis of the two-period change-over design. Biometrics. 1972;28:577–84.

    Article  CAS  PubMed  Google Scholar 

  80. McGowan K, Thomas W, Moran A. Spurious reporting of nocturnal hypoglycemia by CGMS in patients with tightly controlled type 1 diabetes. Diabetes Care. 2002;25:1499–503.

    Article  PubMed  Google Scholar 

  81. Ly TT, Hewitt J, Davey RJ, Lim EM, Davis EA, Jones TW. Improving epinephrine responses in hypoglycemia unawareness with real-time continuous glucose monitoring in adolescents with type 1 diabetes. Diabetes Care. 2011;34:50–2.

    Article  PubMed  Google Scholar 

  82. Choudhary P, Ramasamy S, Green L, Gallen G, Pender S, Brackenridge A, et al. Real-time continuous glucose monitoring significantly reduces severe hypoglycemia in hypoglycemia-unaware patients with type 1 diabetes. Diabetes Care. 2013;36:4160–2.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  83. Ly TT, Nicholas JA, Retterath A, Lim EM, Davis EA, Jones TW. Effect of sensor-augmented insulin pump therapy and automated insulin suspension vs standard insulin pump therapy on hypoglycemia in patients with type 1 diabetes: a randomized clinical trial. JAMA. 2013;310:1240–7.

    Article  CAS  PubMed  Google Scholar 

  84. Pedersen-Bjergaard U, Pramming S, Thorsteinsson B. Recall of severe hypoglycaemia and self-estimated state of awareness in type 1 diabetes. Diabetes Metab Res Rev. 2003;19:232–40.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

We thank Medtronic for providing the study devices and Eli Lilly and Sanofi Aventis for the financial support. Also, we thank all participating patients and primary care centers.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Cornelis A.J. van Beers.

Additional information

Competing interests

This is an investigator initiated trial. Medtronic provided the study devices and accessories. Eli Lilly and Sanofi Aventis provided financial support, but had no role in writing or publication of this manuscript. All authors declare that they have no competing interests.

Authors’ contributions

MD and SK conceived and designed the design. CAJB, ES, PHG, FJS, MHHK participated in the study design. CAJB mainly drafted the initial manuscript and SK, ES, PHG, FJS and MHHK provided advice on the manuscript. All authors read and approved the manuscript.

Additional file

Additional file 1:

Table S1. Overview of visits, telephone consultations and outcome assessments. (XLS 34 kb)

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

van Beers, C.A., Kleijer, S.J., Serné, E.H. et al. Design and rationale of the IN CONTROL trial: the effects of real-time continuous glucose monitoring on glycemia and quality of life in patients with type 1 diabetes mellitus and impaired awareness of hypoglycemia. BMC Endocr Disord 15, 42 (2015). https://doi.org/10.1186/s12902-015-0040-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12902-015-0040-3

Keywords