Table 4 Evaluation of the precision of the automated bioelectrical impedancemeter Tanita® BC-532 to assess the total body fat, compared to the dual-energy x-ray absorptiometry: residual values

From: Body fat assessment in youth with overweight or obesity by an automated bioelectrical impedance analysis device, in comparison with the dual-energy x-ray absorptiometry: a cross sectional study

 

Boys (N = 93)

TBF DEXA (%) =

[0.617 × TBF BIA Tanita® BC-532 (%)] + 26.24

Girls (N = 104)

TBF DEXA (%) =

[0.648 × TBF BIA Tanita® BC-532 (%)] + 22.17

Residuals (%)

Mean ± SD

− 9.92 ± 6.30

−9.36 ± 5.59

Min

−24.20

− 24.00

Max

13.60

15.40

Residuals (absolute value)

Mean ± SD

10.52 ± 5.22

9.96 ± 4.40

Max

24.20

24

  1. TBF DEXA (%): total body fat percentage assessed with dual-energy x-ray absorptiometry
  2. TBF BIA Tanita® BC-532 (%): total body fat percentage assessed with the bioelectrical impedance analyser Tanita® BC-532
  3. Residuals (%) = [TBF BIA BIA Tanita® BC-532 (%)-TBF DEXA (%)]